Building better technology
for the civil engineering industry

Building better technology
for the civil engineering industry

Tel: +44 (0)118 945 3737

Tel: +44 (0)118 945 3737

VJTech Limited Mobile Logo

signin Login register Register Tel: +44 (0)118 945 3737

 
     
Sample Preparation - Rowe Cell

Sample Preparation - Rowe Cell

Oct 08 2019

1.  Introduction

The following items shown in Figure 1 will be required to prepare a sample and place into a VJ Tech Rowe cell.

              

Using the cutter ring, create a test specimen. This can be done either by pushing the ring (cutting edge first) into a block of soil or undisturbed specimen of a larger diameter than the ring; or remoulding a sample into the ring. Use the trimming tool to cut excess material from around the cutter ring and level the specimen (Figure 2).


2.  Placing Sample into Rowe Cell Mould

First turn the body of the Rowe cell upside down and place on top of a bench (Figure 3).


Next place the extension ring over the base of the cell as shown in Figure 4.

Next carefully place the cutter ring with the sample into the extension ring as shown in Figure 5.

Place the dolly onto the specimen (Figure 6), and carefully push the sample out of the cutter ring into the Rowe cell body.

Stop pushing the sample once the sample is located just below to top of the cell body and remove the cutter and extension ring (Figure 7).

Now carefully turn the cell body the correct way up and place onto the base of the cell.

 

 