Preparation for a CRS test in a Multi-Purpose Cell

Preparation for a CRS test in a Multi-Purpose Cell

Jul 25 2019

Constant Rate of Strain test – Check list

The following check list is designed to help a new user complete all the necessary tasks required before starting a CRS test using the Multi-Purpose Cell from VJTech.

The test can be performed with either an ACONS Pro or a TriSCAN Pro load frame. The following guide refers to the procedure followed on an ACONS Pro load frame. The user should also reference the local standard for additional steps and detail on how to undertake certain steps below.

1.     Ensure a supply of de-aired water.

2.     Place the porous disc inside a tray with de-aired water and leave to saturate.

3.     Clean the base of the MP cell and slightly grease O-ring in base groove.

4.     Flush the de-airing block.

5.     Check pore water pressure transducer calibration and recalibrate if necessary. The calibration needs to be performed to the instrument that the sensor is connected to (look at point 17 below).

6.     Flush the Back Pressure and Pore Pressure lines. Leave a small water bubble on the pedestal.

7.     Check Back Pressure line for leaks.

8.     Attach the base pedestal required for CRS tests in the MPC base. Make sure that the two holes at the bottom of the pedestal are aligned with the holes at the bottom of the cell (figure 1). Use the 5 bolts to tighten it up firmly.

 

1.   Weigh the empty cutting ring and prepare the sample (disturbed or undisturbed). Then, weigh the sample & ring and calculate the net weight of the sample. Place any trimmings of the soil in the oven to dry and to determine its moisture content.

2.    Remove the porous disc from the tray and place it onto the base pedestal.

3.     Place the sample on the base pedestal and make sure that the ring is in contact with the O-ring. Then, place on the clamping ring and firmly fix in place using the screws (figure 2 (L)). Place on the top cap (figure 2 (R)).

 

1.     Place on the top of the MP cell while keeping the loading ram up, so that it does not yet come into contact with the top cap. The ram should be held in place using the LSCT bracket. Tighten the three bolts to secure the cell Top Plate tightly (figure 3).

 

1.    Connect the solenoid valve to the Back Pressure line and the Pore Pressure transducer to the Pore Pressure line through the quick-release valves.

2.    Create a junction in the Back Pressure line, placed behind the solenoid valve, to split it between the Cell (CP) and the Back (BP) valves using a T-piece on the pipe (figure 4).

1.     Open the CP valve and close the BP valve. Fill the cell by opening the access to the de-aired water tank through the distribution panel.

2.     Connect the Load Cell and the Displacement transducers to the back of the ACONS Pro instrument, using the designated ports. If a digital displacement transducer is to be used, connect it to the digital port at the back of the ACONS Pro instrument.

3.     Connect the Pore Water Pressure transducer to the External Channel port of the APCH. If the test is conducted using a TriSCAN Pro load frame, the PWP transducer can be connected to one of the spare instrument ports.

4.     Connect the loading ram to the frame’s cross beam (figure 5).

5.     Loosen the LSCT bracket to leaving the loading ram to move freely when the instrument operates. Fix the LSCT in place.

6.     Press Engage on the instrument front panel to engage the loading ram with the sample. The instrument will stop loading as soon as the load changes by the value set up as the Engage Level in the instrument menu. 3 N has been proved to work well in most cases.

7.     Program the test in Clisp studio.

8.     Before starting, make sure that:

a.     The PWP, CP and BP valves on the cell are open;

b.     The release valve for the de-airing block of the PWP sensor is closed;

c.     The sample is properly engaged with a very small load (2-3 N);

d.     BP controller is at a value of approximately 0 kPa.

e.     The LSCT bracket is positioned so that is free to move downwards during the test, without hitting any part at the top of the cell.

 

1.     Start the test

 

Total: 0 Comment(s)

Latest Posts

Sample Preparation - Rowe Cell
10/8/2019 3:26 PM | Peter Gawen (VJ Tech)
Preparation for a CRS test in a Multi-Purpose Cell
7/25/2019 4:41 PM | Peter Gawen (VJ Tech)
Benefits of the Multi-Purpose Frame
6/27/2019 10:26 AM | Peter Gawen (VJ Tech)
Global Wind Day
6/14/2019 10:07 AM | Peter Gawen (VJ Tech)
World Environment Day
6/5/2019 2:53 PM | Peter Gawen (VJ Tech)

Categories


Archives

2019 Oct 12 1
2019 Jul 21 1
2019 Jun 85 3
2019 May 460 1
2018 Mar 3831 1
2017 Oct 17401 1
2017 Aug 310 1
2016 Sep 44 1
2016 Aug 66 1
2016 Jul 7 1
2016 May 29 1
2016 Apr 41 1
2016 Mar 21 1
2016 Feb 115 3
2016 Jan 97 1
2015 Dec 29 1
2015 Nov 97 1
2015 May 192 1
2015 Mar 671 4
2015 Feb 554 1
2014 Nov 1938 8
2014 May 396 6
2014 Apr 397 7
2014 Mar 1124 12
2013 Dec 130 2

Tags