Equipment Maintenance part 4

Apr 30 2014

Equipment Maintenance Part 4 – Attaching the ACONS Earth Lead and Adjusting the Screen Contrast.

These 'Equipment Maintenance' posts are intended to show how to service and perform basic maintenance of different VJ Tech equipment. This post describes how to attach the ACONS earth lead should it prove necessary and how to adjust the screen contrast on the ACONS and other various VJ Tech Instruments. This may be necessary if the screen is too bright or too dark.

Attaching the ACONS Earth Lead

Remove the ACONS Front Panel by unscrewing the four bolts and lay the Front Panel down next to the chassis (there should be no need to remove the existing connectors).

Remove the nut on the bottom corner of the Graphics PCB attached to the Front Panel and then use to secure the eyelet on the supplied earth Lead (blue push-fit end) as shown below.

Remove the bolt from the corner of the Main ACONS PCB and use to secure the other end of the supplied Earth Lead as illustrated.

The resultant connection should be as shown below.

Re-attach the ACONS Front Panel ensuring that no wires are trapped.

Adjusting the Screen Contrast

It may be necessary to adjust the screen contrast on the ACONS (or other VJ Tech Equipment) if the screen is too bright or too dark.

The following table will tell you which keys to use on the ACONS and other VJ Tech Equipment; press the selected key a few times to see any change.

Automatic Pressure Controller, MPX300 & ACONS

Keep finger on [3] key to increase Light. Keep finger on [6] key to decrease Light.

MiniScanner2

Press [Shift] key once, and then keep finger on [Enter] key to increase Light.

Press [Shift] key once, and then keep finger on [Esc] key to decrease Light.

TriSCAN

Keep finger on [Pg Up] key to increase Light.

Keep finger on [Pg Down] key to decrease Light.

ShearSCAN 2, ShearTEST and RingSHEAR

Keep finger on [ ] key to increase Light. Keep finger on [] key to decrease Light.

Total: 0 Comment(s)

Latest Posts

Triaxial Testing - an Introduction
10/30/2017 2:03 PM | Peter Gawen (VJ Tech)
Triaxial Cell Sand Sample Preparation
8/14/2017 9:38 AM | Peter Gawen (VJ Tech)
Clisp Studio Release V2.5.5.2426
10/11/2016 8:57 AM | Prash
Predicting long term performance of Offshore Wind Turbines using Cyclic Simple Shear
9/15/2016 4:04 PM | Prash
First VJ Tech Open Day
7/19/2016 2:06 PM | Prash

Categories


Archives

2018 Mar 2828 1
2018 Feb 587 1
2017 Oct 5156 1
2017 Aug 180 1
2016 Oct 42 1
2016 Sep 27 1
2016 Jul 6 1
2016 May 25 1
2016 Apr 28 1
2016 Mar 14 1
2016 Feb 76 3
2016 Jan 88 1
2015 Dec 27 1
2015 Nov 76 1
2015 May 139 1
2015 Mar 476 4
2015 Feb 383 1
2014 Nov 1679 8
2014 May 289 6
2014 Apr 298 7
2014 Mar 893 12
2013 Dec 111 2

Tags