Equipment Maintenance part 4

Apr 30 2014

Equipment Maintenance Part 4 – Attaching the ACONS Earth Lead and Adjusting the Screen Contrast.

These 'Equipment Maintenance' posts are intended to show how to service and perform basic maintenance of different VJ Tech equipment. This post describes how to attach the ACONS earth lead should it prove necessary and how to adjust the screen contrast on the ACONS and other various VJ Tech Instruments. This may be necessary if the screen is too bright or too dark.

Attaching the ACONS Earth Lead

Remove the ACONS Front Panel by unscrewing the four bolts and lay the Front Panel down next to the chassis (there should be no need to remove the existing connectors).

Remove the nut on the bottom corner of the Graphics PCB attached to the Front Panel and then use to secure the eyelet on the supplied earth Lead (blue push-fit end) as shown below.

Remove the bolt from the corner of the Main ACONS PCB and use to secure the other end of the supplied Earth Lead as illustrated.

The resultant connection should be as shown below.

Re-attach the ACONS Front Panel ensuring that no wires are trapped.

Adjusting the Screen Contrast

It may be necessary to adjust the screen contrast on the ACONS (or other VJ Tech Equipment) if the screen is too bright or too dark.

The following table will tell you which keys to use on the ACONS and other VJ Tech Equipment; press the selected key a few times to see any change.

Automatic Pressure Controller, MPX300 & ACONS

Keep finger on [3] key to increase Light. Keep finger on [6] key to decrease Light.

MiniScanner2

Press [Shift] key once, and then keep finger on [Enter] key to increase Light.

Press [Shift] key once, and then keep finger on [Esc] key to decrease Light.

TriSCAN

Keep finger on [Pg Up] key to increase Light.

Keep finger on [Pg Down] key to decrease Light.

ShearSCAN 2, ShearTEST and RingSHEAR

Keep finger on [ ] key to increase Light. Keep finger on [] key to decrease Light.

Total: 0 Comment(s)

Latest Posts

Introduction to Permeability Testing
9/11/2020 10:23 AM | Irina
Introduction to unsaturated triaxial testing
7/3/2020 2:27 PM | Vasileios
Introduction to Bender Element testing
7/3/2020 2:04 PM | Vasileios
CLISP Studio Release 2.7.0.0
4/23/2020 3:09 PM | Adrian Rose
Sample Assembly for the DSS with Confining Pressure
2/10/2020 10:32 AM | Peter Gawen (VJ Tech)

Categories


Archives

2020 Sep 26 1
2020 Jul 208 2
2020 Apr 80 1
2020 Feb 46 1
2019 Nov 1307 1
2019 Oct 36 1
2019 Jul 73 1
2019 Jun 138 3
2019 May 2971 1
2018 Mar 12486 1
2017 Oct 38815 1
2017 Aug 410 1
2016 Sep 62 1
2016 Aug 87 1
2016 Jul 7 1
2016 May 29 1
2016 Apr 58 1
2016 Mar 31 1
2016 Feb 145 3
2016 Jan 116 1
2015 Dec 32 1
2015 Nov 121 1
2015 May 257 1
2015 Mar 846 4
2015 Feb 699 1
2014 Nov 2275 8
2014 May 476 6
2014 Apr 463 7
2014 Mar 1283 12
2013 Dec 79 1

Tags