Equipment Maintenance part 4

Apr 30 2014

Equipment Maintenance Part 4 – Attaching the ACONS Earth Lead and Adjusting the Screen Contrast.

These 'Equipment Maintenance' posts are intended to show how to service and perform basic maintenance of different VJ Tech equipment. This post describes how to attach the ACONS earth lead should it prove necessary and how to adjust the screen contrast on the ACONS and other various VJ Tech Instruments. This may be necessary if the screen is too bright or too dark.

Attaching the ACONS Earth Lead

Remove the ACONS Front Panel by unscrewing the four bolts and lay the Front Panel down next to the chassis (there should be no need to remove the existing connectors).

Remove the nut on the bottom corner of the Graphics PCB attached to the Front Panel and then use to secure the eyelet on the supplied earth Lead (blue push-fit end) as shown below.

Remove the bolt from the corner of the Main ACONS PCB and use to secure the other end of the supplied Earth Lead as illustrated.

The resultant connection should be as shown below.

Re-attach the ACONS Front Panel ensuring that no wires are trapped.

Adjusting the Screen Contrast

It may be necessary to adjust the screen contrast on the ACONS (or other VJ Tech Equipment) if the screen is too bright or too dark.

The following table will tell you which keys to use on the ACONS and other VJ Tech Equipment; press the selected key a few times to see any change.

Automatic Pressure Controller, MPX300 & ACONS

Keep finger on [3] key to increase Light. Keep finger on [6] key to decrease Light.

MiniScanner2

Press [Shift] key once, and then keep finger on [Enter] key to increase Light.

Press [Shift] key once, and then keep finger on [Esc] key to decrease Light.

TriSCAN

Keep finger on [Pg Up] key to increase Light.

Keep finger on [Pg Down] key to decrease Light.

ShearSCAN 2, ShearTEST and RingSHEAR

Keep finger on [ ] key to increase Light. Keep finger on [] key to decrease Light.

Total: 0 Comment(s)

Latest Posts

On sample transducers - Sample preparation
4/1/2021 1:57 PM | Irina@vjtech.co.uk
Introduction to Direct Simple Shear (DSS) Testing
1/26/2021 9:45 AM | adrian@vjtech.co.uk
CLISP Studio Release 2.10.1.0
12/16/2020 9:52 AM | adrian@vjtech.co.uk
Introduction to Permeability Testing
9/11/2020 10:23 AM | Irina@vjtech.co.uk
Introduction to unsaturated triaxial testing
7/3/2020 2:27 PM | vasileios@vjtech.co.uk

Categories


Archives

2021 Apr 45 1
2021 Jan 781 1
2020 Dec 49 1
2020 Sep 4498 1
2020 Jul 2176 2
2020 Feb 69 1
2019 Nov 2864 1
2019 Oct 47 1
2019 Jul 97 1
2019 Jun 108 1
2019 May 6076 1
2018 Mar 23867 1
2017 Oct 59992 1
2017 Aug 517 1
2016 Sep 73 1
2016 Jul 7 1
2016 May 31 1
2016 Apr 73 1
2016 Feb 167 3
2016 Jan 129 1
2015 Dec 40 1
2015 Nov 155 1
2015 Mar 984 4
2015 Feb 813 1
2014 Nov 2678 8
2014 May 554 6
2014 Apr 532 7
2014 Mar 1113 10
2013 Dec 89 1

Tags