Building better technology
for the civil engineering industry

Building better technology
for the civil engineering industry

Login Login Register 

Search

Tel: +44 (0)118 945 3737

Tel: +44 (0)118 945 3737

VJTech Limited Mobile Logo Search

 Login   Tel: +44 (0)118 945 3737

 
     

DSC2000 Calibration - Pressure Transducer

Mar 30 2015

 

DSC2000 Calibration - Pressure Transducer

A PDF copy of this blog is available. Please find SUP0074 in the support section.


Overview

1)   It is recommended that Transducer Calibration is carried out by qualified personnel at a periodic interval defined by the local Standards.

2)   The recommended method for Dual Point Calibration is to use a certified Reference Device that is at least as accurate as the resolution required for the Transducer to be calibrated.

3)   This Support Document covers dual point calibration in CLISP Studio for Pressure transducers connected to a DSC 2000M for dynamic triaxial testing and a DSC 2000MM for Dynamic simple shear or Dynamic triaxials with dynamic cell pressure control.

4)   It is assumed that:

-          CLISP Studio is running

-          The DSC to which the Transducer is connected has been added in CLISP Studio

-          The DSC 2000 is communicating with Clisp Studio

 

Calibrating a Pressure Transducer

To start the calibration processes open the DSC 2000 Instrument Window in Clisp Studio and select the Analogue Inputs Tab



Important

When calibrating a pressure transducer it is very important that the correct units are used. The transducer units must be set to MPa in Clisp Studio.

To edit the calibration setup it may be required that you press the Edit button and enter the passcode 2603.

 

Calibration Initial Setup

Now add the zero and FSD. The Zero is normally 0 and the FSD is the maximum pressure that the transducer can read. In most cases this is 1 MPa. This can be found on the Calibration certificate supplied with the transducer.

 

 

Lastly make sure Dual Point is selected. Single factor calibrations cannot be used with the DSC 2000. Once this has been completed press the ‘Write’ button (see Figure 4).

There are two ways the calibration can be completed either by using the calibration certificate or using a reference device. 

Calibration using the Calibration Certificate

In Clisp Studio open the advanced tab, the first number we need to enter is the Raw Divisions at Full Scale

This number can be calculated from a value that is found on the calibration certificate; the number required is the Rated Pressure Span Output as is highlighted in figure 3.

This number must be put through the following calculation:

So from the example calibration certificate in figure 3 above you would get the following answer:

100.07 / 10000 = 0.010007

So the following would be entered into the raw divisions at full scale. Once this has been entered press the write button.

To complete the Raw divisions at Zero ensure that the transducer is at is zero pressure (usually when released to Atmospheric pressure). Then press the Zero button.

Calibration using Reference Device (APCH)

It is normal to use an APC(H) as a Pressure Reference Device for calibrating a Pore Water Pressure (PWP) Transducer (APC Calibration is covered in SUP0033).

1)     Attach the Pore Water Pressure transducer to one of the Pore Water Pressure valves on the Triaxial Cell Base (Figure 8) and plug the Din plug into the designated socket on the Logger/Instrument. It is important to connect the Transducer to the correct valve on the Triaxial Cell Base (the two PWP valves are directly opposite each other, in line with the two small holes in the Base Pedestal).

1)     Ensure the APC(H) is correctly initialised and then connect the tubing to the PWP, see Figure 8.

2)     Click on the Pressure Channel Number to display the Channel setup window (Figure 10). In order to enable editing of the Calibration Parameters, the ‘Edit’ button should be selected and the passcode entered.

 

1)     Select the ‘Dual Point’ radio button, set the ‘Zero’ field to the lowest Pressure reading (usually 0) and the ‘FSD’ (Full Scale Deflection) field to the highest reading (in this example 1MPa).

2)     To reset the Pressure Channel, click on the ‘Clear’ button.

3)     To send the input information to the Transducer, click on the ‘Write’ button.

4)     With zero Pressure set at the APC, click on the ‘Zero’ button. The Display will reset to 0 – see Figure 9.

5)     Now set the APC to the Maximum Pressure of the Pressure Transducer (in this example 1MPa i.e. 1000kPa) and click on the ‘Gain’ button. The Display will reset to the FSD value (Figure 10)

NOTE: a Reference point must be used to determine zero and maximum pressure.

6)     When different pressures are applied, the Display will reflect the pressure accordingly.

 

 

 

 

 

 


 

Total: 0 Comment(s)

Latest Posts

Advancing Geotechnical Analysis: The Shift to Automation
11/10/2023 9:32 AM | peter.gawan@vjtech.co.uk
OCR: Understanding Over Consolidation Ratio in Geotechnical Design
7/14/2023 3:59 PM | peter.gawan@vjtech.co.uk
Understanding Dynamic Simple Shear Testing: DSS vs DSS-C
5/26/2023 3:00 PM | peter.gawan@vjtech.co.uk
Comparisons between Bromhead and Bishop Ring Shear Testing
4/17/2023 4:00 PM | peter.gawan@vjtech.co.uk
Clisp Studio Major Update Announcement
8/10/2022 4:51 PM | Irina@vjtech.co.uk

Categories


Archives

2023 Nov 7 1
2023 Jul 503 1
2023 May 130 1
2023 Apr 335 1
2022 Aug 308 1
2021 Dec 748 1
2021 Aug 272 1
2021 Apr 124 1
2021 Jan 8760 1
2020 Dec 127 1
2020 Sep 17767 1
2020 Jul 10485 2
2020 Feb 163 1
2019 Nov 8021 1
2019 Oct 109 1
2019 Jul 244 1
2019 Jun 146 1
2019 May 13245 1
2018 Mar 50161 1
2017 Oct 100000 1
2017 Aug 950 1
2016 Sep 245 1
2016 Jul 11 1
2016 May 33 1
2016 Apr 117 1
2016 Feb 256 3
2016 Jan 199 1
2015 Dec 56 1
2015 Nov 231 1
2015 Mar 1550 4
2015 Feb 1146 1
2014 Nov 3788 8
2014 May 809 6
2014 Apr 804 7
2014 Mar 1454 10
2013 Dec 117 1

Tags