De-Aired Water Tank Installation

De-Aired Water Tank Installation

Feb 20 2015

De-Aired Water Tank Installation

This blog is available in a pdf format. Please see Support document SUP0077.

Introduction

The VJ Tech de-airing tank is simple to use, low cost and suitable for most modern laboratories. It can easily be wall mounted by the welded lugs in the top and base plates.

 

Setting up 

The water tank should be mounted securely as high as possible.

The tank is filled via the water inlet marked on figure 2 below; it can be connected to the water supply with standard 15 mm copper pipe or flexible hose. The water is sprayed into the tank through a fine nozzle marked on figure 2 which assists in creating bubbles to release the air from the water.

The vacuum pump, which is connected to vacuum inlet marked on figure 2 by flexible hose, will suck the air bubbles out of the water.

Ensure that the vacuum pump is filled with oil and the filter is fitted correctly.

The de-aired water is released to the distribution panel via the 8 mm push fitting, which is supplied in the base plate as standard.

It is recommended that a valve is fitted in line with the pipe going from the vacuum pump to the de-airing tank so that the vacuum can be released from the tank after de-airing.

A general layout of the de-airing system can be seen in figure 1.

Figure 1 - Connections to De-aired water tank


 

Filling the Vacuum Pump with Oil

The Vacuum pump must be filled with oil before it can be switched on.

 

 

 

Figure 2 - Vacuum Pump

1)     Remove the cover (unlatch the clips and turn clockwise) see Figure 3

Figure 3 - Removing top cover

2)     Undo locking screw for muffler (anti-clockwise), see Figure 4

 

Figure 4 - Removing muffler


3) Fill the Vacuum pump with Oil (use oil supplied by VJ Tech) . NOTE: Do not fill more than the MAX level shown on the sight window.

 

 

Figure 5 - Fill Vacuum Pump with Oil 


 

Using the Tank

Before filling the tank, make sure that the de-airing tank is isolated from the distribution panels of the triaxial system. Close the valve at the entry point into the distribution panel.

Turn on the main water supply to fill the tank with water via the spray nozzle (be careful not to overfill the tank, fill to a maximum level of ¾). Turn the water off and now simply turn on the vacuum pump to remove the air from the water.

As the air is being sucked out of the tank you will notice air bubbles rising up the side of the water in the tank.

Whilst the vacuum is being applied to the water, the tank must not be opened to the distribution panel. If a sudden vacuum is applied to a triaxial system by opening the de-airing tank to the distribution panel then it is possible that the pressure controllers and pressure transducers could be damaged. The tank must never be pressurised.

When air bubbles are no longer visible on the side of the de-airing tank switch off the vacuum pump.

Before the water can be used the vacuum in the tank must be released. Failure to release the vaccum will stop water running freely to your triaxial system and could possibly damage components of your triaxial system.

When additional water is introduced into the tank the vacuum pump must be switched on to de-air the new water.

 

Maintenance

As required, the tank may be periodically disassembled and cleaned. Simply unscrew the 4 cap nuts and carefully remove the top plate then clean the cylinder with a mild detergent. When replacing the cylinder ensure that it is securely located on the o – ring (a lubricant may be used).

 

Maintenance for a vacuum pump supplied by VJ tech can be found in the relevant user guide, please contact service@vjtech.co.uk to obtain a copy.

 


Latest Posts

Saturation Procedure for HAE Ceramic Discs in SWCC Systems
8/26/2021 4:34 PM | Irina@vjtech.co.uk
On sample transducers - Sample preparation
4/1/2021 1:57 PM | Irina@vjtech.co.uk
Introduction to Direct Simple Shear (DSS) Testing
1/26/2021 9:45 AM | adrian@vjtech.co.uk
CLISP Studio Release 2.10.1.0
12/16/2020 9:52 AM | adrian@vjtech.co.uk
Introduction to Permeability Testing
9/11/2020 10:23 AM | Irina@vjtech.co.uk

Categories


Archives

2021 Aug 5 1
2021 Apr 47 1
2021 Jan 1130 1
2020 Dec 55 1
2020 Sep 5288 1
2020 Jul 2627 2
2020 Feb 74 1
2019 Nov 3177 1
2019 Oct 50 1
2019 Jul 109 1
2019 Jun 109 1
2019 May 6551 1
2018 Mar 25197 1
2017 Oct 63140 1
2017 Aug 531 1
2016 Sep 164 1
2016 Jul 7 1
2016 May 31 1
2016 Apr 73 1
2016 Feb 177 3
2016 Jan 136 1
2015 Dec 40 1
2015 Nov 157 1
2015 Mar 1011 4
2015 Feb 833 1
2014 Nov 2745 8
2014 May 563 6
2014 Apr 547 7
2014 Mar 1136 10
2013 Dec 90 1

Tags