Building better technology
for the civil engineering industry

Building better technology
for the civil engineering industry

Tel: +44 (0)118 945 3737

Tel: +44 (0)118 945 3737

VJTech Limited Mobile Logo

 Login  Register Tel: +44 (0)118 945 3737

 
     
Triaxial Cell Sand Sample Preparation

Triaxial Cell Sand Sample Preparation

Aug 14 2017

Triaxial Cell Sand Sample Preparation

1.     This blog describes how to setup a remoulded sand sample for a triaxial test in a standard triaxial cell using VJ Tech’s updated sample former with attachment for a suction device.

2.     In addition to a load frame and cell, the equipment in Figure 1 is also required.


 3.    Ensure the Back Pressure line and base pedestal are flushed with de-aired water and are free from leaks. Clean the base of the cell.

4.     Place the first porous disc onto the base pedestal of the cell (Figure 2 ).

5.     Next place the membrane over the base pedestal (Figure 3 ).

   UsUsing the ‘O’ Ring tool place two ‘O’ Rings over the membrane to secure it on the base pedestal (Figure 4 ).

7.     Take the silicone grease and sample former. Remove the clamp (see Figure 5 ) from the former so the two halves can be separated.

 

 

8.     Along the inside edges of the sample former (shown in Figure 6 ) smear a small amount of silicone grease (other hydrocarbon based greases can damage the membranes).

9.     Place the two halves of the former over the base pedestal taking care not to trap the membrane between them Figure 7 (L).

10.  Then place the former clamp over the former and once in place tighten using the locking screw Figure 7 (R).


Fold membrane over sample and add suction device (Fig 8).



12.  Apply suction using the suction device so that the membrane is stretched tight against the inside of the former (Figure 9 ).

13.  The sand can now be poured into the former to create the required density of sample. The sand can be tamped down using the sample tamper and by gently tapping the side of the former (Figure 10 ).

.  Once the sand has filled the former and has been levelled place the final porous disc on top of the sample (Figure 11 ).

15.  Place an ‘O’ Ring over the top cap and place the top cap on the top porous disc (Figure 12 ).

16.  Fold up the excess membrane over the top cap. Using the ‘O’ Ring tool place the ‘O’ Ring over the membrane and top cap (Figure 13 ).

.  Remove the suction device from the former. A small vacuum will now need to be applied to the sample. The easiest way to do this is to attach the suction device (used for stretching the membrane) to the valve on the base of the triaxial cell that is on the opposite side to the pore pressure valve connection (Figure 14 ). Then create suction using the suction device, this may need to be repeated a few times to remove a large amounts of air from large samples. Once the air is removed and the suction holds shut the valve that the suction device is connected to. Using this device typically applies a small negative pressure of approximately -15 to -20 kPa. By applying a small negative pressure to the sample the sample shouldn’t collapse once the former is removed. Other methods can be used but are not detailed in this document.

 

18.  Carefully undo the sample former clamp and remove it. Then remove the two halves of the sample former again taking care not to damage the sample or membrane. Carefully fold down the excess membrane. Clean any excess sand way from the base of the triaxial cell (Figure 15 ).

 

The triaxial cell can then be assembled (take care not to damage the sample when lifting on the top half of the cell) and then filled with water for testing.

 

Tips:

 

Before attempting saturation the following can be attempted to reduce the time and pressures required when saturating a sand sample:

 

-       Flush the sample with de-aired water. This should be done from base to top. To do this, increase the cell pressure to approximately 30 kPa. Remove the suction device to release the vacuum on the sample. Connect a line to the same valve as used by the suction device from the de-aired water tank. Open the Back Pressure valve on the cell and allow water to flow up the sample. Do this until no air is seen coming out of the Back Pressure line. After this flush, set the sample up for testing. Continue with the normal saturation procedures.

-       After flushing the sample (as described above) close the valve to the de- aired water supply and the Back Pressure valve. Increase the Cell Pressure to a pressure greater than 300 kPa. Leave the sample like this for 12 hours (overnight). After this, flush the sample again with clean de-aired water. Then continue with the normal saturation process.

-       The sample can be flushed with CO2. CO2 will dissolve more easily into water than air so reducing the saturation time.


Latest Posts

Triaxial Testing - an Introduction
10/30/2017 2:03 PM | Peter Gawen (VJ Tech)
Triaxial Cell Sand Sample Preparation
8/14/2017 9:38 AM | Peter Gawen (VJ Tech)
Clisp Studio Release V2.5.5.2426
10/11/2016 8:57 AM | Prash
Predicting long term performance of Offshore Wind Turbines using Cyclic Simple Shear
9/15/2016 4:04 PM | Prash
First VJ Tech Open Day
7/19/2016 2:06 PM | Prash

Categories


Archives

2017 Oct 7 1
2017 Aug 42 1
2016 Oct 24 1
2016 Sep 20 1
2016 Jul 4 1
2016 May 24 1
2016 Apr 16 1
2016 Mar 11 1
2016 Feb 58 3
2016 Jan 61 1
2015 Dec 23 1
2015 Nov 52 1
2015 May 81 1
2015 Mar 316 4
2015 Feb 238 1
2014 Nov 1124 8
2014 May 177 6
2014 Apr 203 7
2014 Mar 656 12
2013 Dec 96 2

Tags